Technical Specifications
The area of application for the device includes clinics, hospitals, ambulances, and field conditions.
The device is designed for artificial ventilation of the lungs with active inhalation and passive exhalation in any emergency situations involving acute respiratory failure.
The inclusion of a portable foot-operated suction device in the equipment allows for the suction of fluids from a person’s airways.
The volume of air delivered per complete compression of the bag, cm³ (l) not less than | 1200 (1,2) |
The leakage of the breathing bag assembled with the breathing and suction valves at a pressure in the internal cavity of the bag of 3+0.3 kPa (300+30 mm water column), l/min, not more than. | 2 |
Pressure loss in the passive exhalation line and the additional inhalation valve at a constant flow of 25 l/min, kPa (mm water column), not more than |
0,2 (20) |
Time for complete deflation of the bag, s, not more than | 1 |
The device is operational under the influence of: | |
Air temperature, °C | from +5 to +40 |
Relative humidity of the air at an air temperature of (25±2) °C, % Overall dimensions, mm, not more than | 98 |
Length | 390 |
Width | 150 |
Height Weight, kg, not more than. Average service life, years, not less than | 305 |