It is one of the components for the ADR-1 device and is supplied separately upon the customer’s request.
The device is intended for short-term manual artificial ventilation of the lungs in cases of acute respiratory failure.
The device is used in emergency situations, intensive care units, and field conditions.
The volume of air delivered per complete compression of the bag, cm³ (l) not less than | 1200 (1,2) |
The leakage of the breathing bag assembled with the breathing and suction valves at a pressure in the internal cavity of the bag of 3+0.3 kPa (300+30 mm water column), l/min, not more than | 2 |
Pressure loss in the passive exhalation line and the additional inhalation valve at a constant flow of 25 l/min, kPa (mm water column), not more than | 0,2 (20)
|
Time for complete deflation of the bag, s, not more than | 1 |
The device is operational under the influence of: | |
Air temperature, °C |
from +5 to +40 |
Relative humidity of the air at an air temperature of (25±2) °C, % Overall dimensions, mm, not more than: | 98 |
Length | 390 |
Width | 150 |
Height | 305 |
Weight, kg | 2,5 |
Average service life, years, not less than | 4 |